Test Code LAB10594 GI PANEL

Performing Laboratory

 

Mohawk Valley Health System Laboratory

Day(s) Performed

Performed: Daily

Routine TAT: Same Day

Available as a STAT test

Intended Use

Rapid detection of gastrointestinal infections caused by:

-Campylobacter species (Campylobacter jejuni/Campylobacter coli/Campylobacter upsaliensis)

-Plesiomonas shigelloides

-Salmonella species

-Yersinia species

-Enteropathogenic E coli (EPEC)

-Enterotoxigenic E coli (ETEC)

-Shiga toxin

-E coli O157

-Shigella/Enteroinvasive E coli (EIEC)

-Cryptosporidium species

-Cyclospora cayetanensis

-Entamoeba histolytica

-Giardia

-Adenovirus F 40/41

-Astrovirus

-Norovirus GI/GII

-Rotavirus A

 

This test is not recommended as a test of cure.

Ordering Guidance

It is not recommended that the following tests be concomitantly ordered if this test is ordered:

-ROTA / Rotavirus Antigen, Feces

-LADV / Adenovirus, Molecular Detection, PCR, Varies

-GIAR / Giardia Antigen, Feces

-CRYPS / Cryptosporidium Antigen, Feces

-CYCL / Cyclospora Stain, Feces

-STL / Enteric Pathogens Culture, Feces

-CAMPC / Campylobacter Culture, Feces

-SHIGC / Shigella Culture, Feces

-SALMC / Salmonella Culture, Feces

-YERSC / Yersinia Culture, Feces

-E157C / Escherichia coli O157:H7 Culture, Feces

-STFRP / Shiga Toxin, Molecular Detection, PCR, Feces

-LNORO / Norovirus PCR, Molecular Detection, Feces

Specimen Type

Stool

Collection Container

ParaPak, Sterile container, Sterile cup

Specimen Collection and Handling

Acceptable specimen Type: Stool Samples re-suspended in transport medium (Para-Pak C&S (Meridian Bioscience))

 

Preserved specimen may be stored at:

Room temperature up to 4 days at 15–25°C
Refrigerated up to 4 days at 2–8°C

 

Unpreserved specimen must be received in lab or placed into ParaPak transport media vial within 2 hours of collection
 

Unacceptable Specimens: Swab specimens, unpreserved specimens over 2 hours, Specimens in incorrect preservative vial (formalin (eg, Sodium Acetate-Acetic Acid Formalin fixative [SAF]; PolyVinyl Alcohol fixative [PVA]; EcoFix preservative) or Unapproved commercial transport media.

 

Testing should not be performed on formed stool specimens. Formed stool spcimens may be rejected by Laboratory.
 

 

Notes

1. Results from the QIAstat-Dx Gastrointestinal Panel 2 are not intended to be used as the sole basis for diagnosis, treatment or other patient management decisions. All assay results should be used and interpreted by a trained healthcare professional in the context of a full clinical evaluation, laboratory and epidemiological findings, as an aid in the diagnosis of gastrointestinal infection. Positive results do not rule out co-infection with organisms not included in the QIAstat-Dx Gastrointestinal Panel 2. The agent detected may not be the definitive cause of the disease. 
2. Genetic virulence markers associated with diarrheagenic E. coli/Shigella pathotypes are often carried on mobile genetic elements (MGEs) that can be transferred horizontally between different strains; therefore, “Detected“ results for multiple diarrheagenic E. coli/Shigella may be due to co-infection with multiple pathotypes or, less frequently, may be due to the presence of a single organism containing genes characteristic of multiple pathotypes. 
3. The QIAstat-Dx Gastrointestinal Panel 2 detects Enteropathogenic E. coli (EPEC) in Para-Pak C&S collected samples through targeting of the eae gene, which encodes the adhesin intimin. As some Shiga-like toxin-producing E. coli (STEC) also carry eae (in particular, strains identified as enterohemorrhagic E. coli; EHEC), the QIAstat-Dx Gastrointestinal Panel 2 cannot distinguish between STEC containing eae and a co-infection of EPEC and STEC. Therefore, the EPEC result is not applicable (N/A) and not reported for specimens in which STEC has also been detected. In rare cases, STEC may be reported as EPEC when an STEC carrying eae (EHEC) is present in a specimen below the LoD of the STEC oligonucleotide design (s). Rare instances of other organisms carrying eae have been documented; e.g., Escherichia albertii and Shigella boydii 
4. E. coli O157 result is only reported as specific serogroup identification in association with STEC stx1/stx2 in Para-Pak C&S collected samples. 
5. The performance of the QIAstat-Dx Gastrointestinal Panel 2 has not been established in individuals who received Rotavirus A vaccine. Recent oral administration of a Rotavirus A vaccine may cause positive results for Rotavirus A if the virus is passed in the stool. 
6. The performance of this test has not been established for monitoring treatment of infection with any of the panel organisms. 
7. Virus, bacteria, and parasite nucleic acid may persist in vivo independently of organism viability. Additionally, some organisms may be carried asymptomatically. Detection of organism targets does not imply that the corresponding organisms are infectious or are the causative agents for clinical symptoms